亞洲光學設備公司
電話: (852) 3188 5568 傳真: (852) 3106 2266  QQ: 2377671905
電郵: info@opticaldevice.asia 網址: www.opticaldevice.asia
-
地址: 香港 九龍 新蒲崗 大有街 31號 善美工業大廈 7字樓 703B室
Canon TX-F Full Auto Tonometer
With a non-contact Tonometer, an air puff nozzle built-in the center of objective lens is placed very close to the surface of the patient’s eye. A puff of air is then blown at center of the cornea, and an instrument calculates pressure from the change in the light reflected off the cornea. One problem with this method is that measurement errors can easily occur if the air puff nozzle is not placed in exactly the right position. The patient’s eye often moves, which necessitates special adeptness on the part of the operator due to the difficulty of aligning the machine manually according to the eye movement.
Specifications
Type | Non-contact type | |
Measurable range |
0 to 60 mmHg Includes automatic shifting between 30 / 60 mmHg. |
|
Measuring increment | 1 mmHg | |
Operation distance | 11.3 mm from the front of nozzle | |
Observation range | Approx. 15 x 12 mm | |
Alignment | Full Auto / Auto / Manual |
Memory | Max. 10 measurements for each eye |
Data output | RS232C |
Power-saving system | Available |
Display | 5-inch monochromeCRT monitor |
Printer | Thermal line printer |
Power supply | AC100 – 240V, 50 / 60Hz, 0.4 – 0.8A |
Power consumption | Approx. 80VA |
Operating range |
Front / back: 40 mm Left / right: 90 mm Up / down: 30mm
|
Dimensions (W x D x H) | 280 x 520 x 495 mm (11.0 x 20.5 x 19.5 in.) |
Weight | 20.5 kg (45.2 lbs.) |